Why Is My Belmont 071 Not Working?

Why-Is-My-Belmont-071-Not-Working ABCDentalworks

Why Is My Belmont 071 Not Working?

Below are the 3 fast checks techs look at first for the Belmont 071 intraoral X-ray.

1) Image / Output Issues

  • Symptoms: blurry or over/underexposed images, artifacting, cutoff edges.

  • Likely causes: tube head wear (oil leak/overheating), mis-calibration of exposure time, cone/collimator misalignment, incorrect sensor settings.

  • Quick checks: confirm patient/sensor positioning and exposure settings; inspect the tube head for heat/oil; verify collimator alignment. Stop use if oil is present and schedule evaluation/replacement.

2) Control / Timer & Power

  • Symptoms: no power, breaker trips, error beeps, won’t fire, inconsistent timing.

  • Likely causes: blown fuse, bad hand-switch or cable, failing timer/control board, transformer or line-voltage issues, loose ground.

  • Quick checks: verify outlet/line voltage, check fuses, inspect hand-switch cable/connectors; if timing is unstable, a board/transformer service may be needed.

3) Arm Drift & Mechanics

  • Symptoms: head sags or won’t hold position; rattling joints.

  • Likely causes: worn friction springs/washers, dried lubrication, loose wall/stand mount.

  • Quick checks: adjust friction, replace springs/washers, lubricate joints, tighten mounting hardware.

⚠️ Safety: X-ray devices must be serviced by qualified personnel. Power down, lock-out/tag-out, and follow state radiation rules.

Need help fast? Call ABC DentalWorks at 727-388-4223 for Belmont 071 diagnostics, parts, and repair.